4.8 Article

Selection of the implant and coating materials for optimized performance by means of nanoindentation

期刊

ACTA BIOMATERIALIA
卷 7, 期 2, 页码 874-881

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.actbio.2010.09.023

关键词

Nanoindentation; Substrate effects; Single droplet; Hydroxyapatite; Thermal spray

资金

  1. ARC [LE0668256]

向作者/读者索取更多资源

Mechanical compatibility between a coating and a substrate is important for the longevity of implant materials. While previous studies have utilized the entire coating for analysis of mechanical compatibility of the surface, this study focuses on the nanoindentation of a uniformly thermally sprayed splat. Hydroxyapatite was thermally sprayed to create a homogeneous deposit density, as confirmed by microRaman spectroscopy, of amorphous calcium phosphate. Substrates were commercially pure Ti, Ti-6Al-4V, Co-Cr alloy and stainless steel. Nanoindentation revealed that splats deposited on the different metals have similar hardness and elastic modulus values of 4.2 +/- 0.2 GPa and 80 +/- 3 GPa, respectively. The mechanical properties were affected by the substrate type more than residual stresses, which were found to be low. It is recommended that amorphous calcium phosphate is annealed to relieve the quenching stress or that appropriate temperature histories are chosen to relax the stress created in cooling the coating assembly. (C) 2010 Acta Materialia Inc. Published by Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据