4.8 Article

Picomolar detection of protease using peptide/single walled carbon nanotube/gold nanoparticle-modified electrode

期刊

ACS NANO
卷 2, 期 5, 页码 1051-1057

出版社

AMER CHEMICAL SOC
DOI: 10.1021/nn8000774

关键词

protease; ferrocene-pepstatin inhibitor; carbon nanotubes; gold nanoparticles; biosensor

向作者/读者索取更多资源

Picomolar electrochemical detection of human immunodeficiency virus type-1 protease (HIV-1 PR) using ferrocene (Fc)-pepstatin-modified surfaces has been presented. Gold electrode surface was modified with gold nanoparticles (AuNP) or thiolated single walled carbon nanotubes/gold nanciparticles (SWCNT/AuNP). Thiol-terminated Fc-pepstatin was then self-assembled on such surfaces as confirmed by Raman spectroscopy and scanning electron microscope. The interaction between the Fc-pepstatin-modified substrates and HIV-1 PR was studied by cyclic voltammetry and electrochemical impedance spectroscopy. Both electrode materials showed enhanced electrochemical responses to increasing concentrations of HIV-1 PR with shifting to higher potentials as well as decrease in the overall signal intensity. However, the sensing electrode modified with thiolated SWCNTs/AuNPs showed remarkable detection sensitivity with an estimated detection limit of 0.8 pM.