4.8 Article

Metal (Au, Pt) Nanoparticle-Latex Nanocomposites as Probes for Immunochromatographic Test Strips with Enhanced Sensitivity

期刊

ACS APPLIED MATERIALS & INTERFACES
卷 10, 期 38, 页码 31977-31987

出版社

AMER CHEMICAL SOC
DOI: 10.1021/acsami.8b11745

关键词

immunochromatographic test; lateral flow test; immunoassay; influenza virus; gold nanoparticles; platinum nanoparticles; localized surface plasmon resonance

资金

  1. Nanotechnology Platform Program of the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan

向作者/读者索取更多资源

The development of a sensitive and rapid diagnostic test for early detection of infectious viruses is urgently required to defend against pandemic and infectious diseases including seasonal influenza. In this study, we developed noble metal (Au, Pt) nanoparticle-latex nanocomposite particles for use as probes for immunochromatographic test (ICT) strips. The nanocomposite particles were conjugated with monoclonal antibody (mAb) to detect an influenza A (H1N1) antigen. For comparison, Au nanoparticles conjugated with mAb were also prepared. The lowest detectable concentrations of the influenza A antigen were found to be 6.25 x 10(-3) and 2.5 X 10(-2) HAU/mL for Au nanoparticle-latex and Pt nanoparticle-latex nanocomposite particles, respectively, whereas it was 4.0 X 10(-1) HAU/mL for Au nanoparticles. These results clearly demonstrated that the nanocomposite probes were more sensitive than conventional nanoparticle-based probes for ICT. To expand the versatility of the nanocomposite probes, the surfaces of the probes were functionalized with biotinylated proteins to enable modification of their surfaces with desired biotinylated antibodies through biotin-avidin binding.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据