4.8 Article

Electrochemical Glucose Sensors Enhanced by Methyl Viologen and Vertically Aligned Carbon Nanotube Channels

期刊

ACS APPLIED MATERIALS & INTERFACES
卷 10, 期 34, 页码 28351-28360

出版社

AMER CHEMICAL SOC
DOI: 10.1021/acsami.8b08997

关键词

electrochemical sensor; glucose; methyl viologen; vertically aligned carbon nanotubes (VACNTs); microchannel; flow injection analysis (FIA)

资金

  1. Department of Mechanical Engineering at Brigham Young University
  2. College of Engineering and Technology at Brigham Young University

向作者/读者索取更多资源

Freestanding, vertically aligned carbon nano tubes (VACNTs) were patterned into 16 mu m diameter microchannel arrays for flow-through electrochemical glucose sensing. Non-enzymatic sensing of glucose was achieved by the chemical reaction of glucose with methyl viologen (MV) at an elevated temperature and pH (0.1 M NaOH), followed by the electrochemical reaction of reduced-MV with the VACNT surface. The MV sensor required no functionalization (including no metal) and was able to produce on average 3.4 electrons per glucose molecule. The current density of the MV sensor was linear with both flow rate and glucose concentration. Challenges with interference chemicals were mitigated by operating at a low potential of -0.2 V vs Ag/AgCl. As a comparison, enzymatic VACNT sensors with platinum nano-urchins were functionalized with glucose oxidase by covalent binding (1-ethyl-3-(-3-dimethylaminopropyl)carbodiimide/N-hydroxysuccinimide) or by polymer entrapment [poly(3,4-ethylene-dioxythiophene)] and operated in phosphate buffered saline. With normalization by the overall cross-sectional area of the flow (0.713 cm(2)), the sensitivity of the MV, enzyme-in solution, and covalent sensors were 45.93, 18.77, and 1.815 mA cm(-2) mM(-1), respectively. Corresponding limits of detection were 100, 194, and 311 nM glucose. The linear sensing ranges for the sensors were 250 nM to 200 mu M glucose for the MV sensor, 500 nM to 200 mu M glucose for the enzyme-in-solution sensor, and 1 mu M to 6 mM glucose for the covalent sensor. The flow cell and sensor cross-sectional area were scaled down (0.020 cm(2)) to enable detection from 200 mu L of glucose with MV by flow injection analysis. The sensitivity of the small MV sensor was 5.002 mA cm(-2) mM(-1), with a limit of detection of 360 nM glucose and a linear range up to at least 150 mu M glucose. The small MV sensor has the potential to measure glucose levels found in 200 mu L of saliva.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据