3.8 Article

Whole blood viscosity issue VIII: Comparison of extrapolation method with diagnostic digital viscometer

期刊

出版社

WOLTERS KLUWER MEDKNOW PUBLICATIONS
DOI: 10.4297/najms.2011.3333

关键词

Digital viscometer; extrapolation chart; whole blood viscosity

向作者/读者索取更多资源

Background: The first issue of this series proposed extrapolation chart with conventional reference range and suggested comparison of results with other methods. Aim : This work sets out to compare interpretative results from the extrapolation method with those from a digital viscometer method. Materials and Methods: Five cases in our archived clinical pathology database that were specifically tested for whole blood viscosity by the digital method, and had results for haematocrit and serum proteins were pooled. The values of haematocrit and serum proteins were used to derive extrapolated values. The interpretative results of the extrapolation method were compared with those of digital viscometer-based clinical reports. Non-Newtonian fluids such as whole blood have different viscosities at different shear rates. Comparative statement can only be based on interpreted outcome. Results: Two-fifth absolute concordance and one-fifth discordance is observed between extrapolation and viscometer-based clinical reports. The discordance is a case of hyperviscosity in the presence of neither hyperproteinaemia nor polycythemia. Conclusion: The extrapolation method may underestimate whole blood viscosity in some patients when compared with digital viscometer, which in turn may suggest hyperviscosity that cannot be explained by hyperproteinaemia or polycythemia concepts. The impact of oxidative stress is highlighted.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据