4.1 Article

Near-infrared autofluorescence imaging to detect parathyroid glands in thyroid surgery

期刊

出版社

ROYAL COLL SURGEONS ENGLAND
DOI: 10.1308/rcsann.2017.0102

关键词

Parathyroid glands; Autofluorescence; Near infrared light; Thyroidectomy

类别

向作者/读者索取更多资源

OBJECTIVE To identify and save parathyroid glands during thyroidectomy by displaying their autofluorescence. METHODS Autofluorescence imaging was carried out during thyroidectomy with and without central lymph node dissection. After visual recognition by the surgeon, the parathyroid glands and the surrounding tissue were exposed to near-infrared light with a wavelength of 690-770 nm using a modified Karl Storz near infrared/indocyanine green endoscopic system. Parathyroid tissue was expected to show near infrared autofluorescence at 820 nm, captured in the blue channel of the camera. RESULTS We investigated 41 parathyroid glands from 20 patients; 37 glands were identified correctly based on near-infrared autofluorescence. Neither lymph nodes nor thyroid revealed substantial autofluorescence and nor did adipose tissue. CONCLUSIONS Parathyroid tissue is characterised by showing autofluorescence in the near-infrared spectrum. This effect can be used to identify and preserve parathyroid glands during thyroidectomy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据