4.5 Review

Shear wave elastography safety in fetus: A quantitative health risk assessment

期刊

DIAGNOSTIC AND INTERVENTIONAL IMAGING
卷 99, 期 9, 页码 519-524

出版社

ELSEVIER MASSON, CORPORATION OFFICE
DOI: 10.1016/j.diii.2018.04.013

关键词

Elastography; Shear waves; Evaluation risk; Ultrasound; Health risk assessment

向作者/读者索取更多资源

Purpose: To identify the effects of shear wave elastography in the fetus for evaluation in widespread use. Materials and methods: The Health Risk Assessment method proposed by the National Research Council was used with literature to evaluate the safety of shear wave elastography for the fetus regarding its potential effects in human tissues. Results: The experimental and epidemiologic data from 25 articles showed that shear wave elastography maintained the same thermal effect as pulsed Doppler ultrasound already authorized in obstetrics, and that cavitation effect on fetal tissue is improbable. Nonetheless, the vibratory character of shear waves could induce displacement of fetal tissue while potential effects of very short duration energy peaks of the radiation force focused wave front remain unknown. Conclusion: The actual knowledge does not provide enough information to assess the effects of shear wave elastography on fetal tissues, thus these points have to be explored by further experimental studies. (C) 2018 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据