期刊
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
卷 62, 期 1, 页码 221-227出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2014.2379941
关键词
Charge preamplifier; silicon drift detectors; X-ray spectroscopy
资金
- ESA-ESTEC [4000102940/11/NL/NR]
- Italian INFN (Kaonnis experiment)
In this work we present the results of the experimental characterization of Silicon Drift Detectors (SDDs) readout by CUBE preamplifiers for X-ray spectroscopy measurements. One specific goal of the work was to characterize SDDs of different sizes cooled at low temperature in view of their use in the upgrade of the SIDDHARTA nuclear physics experiment. Beside the target application, the results of this work are also of interest for a more extended use of the SDDs in other X-ray spectroscopy applications. The SDDs have been designed as single units with square shape of different areas, 64 mm(2) (8 mm x 8 mm) and 144 mm(2) (12 mm x 12 mm ), and also as monolithic array 3 x 3 of elements of the 8 mm x 8 mm unit (total area 26 mm x 26 mm). The read-out of the SDDs is based on a CMOS (Complementary Metal Oxide Semiconductor) preamplifier (CUBE) both for the single unit and for the 3 x 3 array. For the readout of the array, an Application Specific Integrated Circuit (ASIC) has been used. An energy resolution better than 124 eV at the Mn - K alpha line has been measured with a 64 mm(2) SDD cooled at the temperature of 50 K. The energy resolution remains good (< 130 eV) also at short shaping time (250 ns) thanks to the noise feature of the CUBE preamplifier. Results of measurements on SDDs of different format and also on arrays of SDDs are presented in this work.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据