4.6 Article

Comparison of Smartphone Ophthalmoscopy With Slit-Lamp Biomicroscopy for Grading Diabetic Retinopathy

期刊

AMERICAN JOURNAL OF OPHTHALMOLOGY
卷 159, 期 2, 页码 360-364

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ajo.2014.11.008

关键词

-

向作者/读者索取更多资源

PURPOSE: To assess the accuracy and reliability of smartphone ophthalmoscopy, we compared the ability of a smartphone ophthalmoscope with that of a slit-lamp biomicroscope to grade diabetic retinopathy (DR) in patients with diabetes mellitus (DM). DESIGN: Clinical-based, prospective, comparative instrument study. METHODS: This comparative clinical study was performed in 120 outpatients (240 eyes) with type 1 or type 2 DM. After pupil dilation, the patients underwent smartphone ophthalmoscopy with the D-Eye device, followed by dilated retinal slit-lamp examination, to grade DR according to a 5-step scale. RESULTS: Overall exact agreement between the 2 methods was observed in 204 of 240 eyes (85%) (simple K = 0.78; CI 0.71-0.84) and agreement within 1 step was observed in 232 eyes (96.7%). Compared to biomicroscopy, the sensitivity and specificity of smartphone ophthalmoscopy for the detection of clinically significant macular edema were 81% and 98%, respectively. Smart-phone ophthalmoscopy and biomicroscopy could not be used to examine the fundus and grade DR in 9 eyes (3.75%) and 4 eyes (1.7%), respectively, because of cataract and/or small pupil diameter. CONCLUSION: Smartphone ophthalmoscopy showed considerable agreement with dilated retinal biomicroscopy for the grading of DR. The portability, affordability, and connectivity of a smartphone ophthalmoscope make smartphone ophthalmoscopy a promising technique for community screening programs. (C) 2015 by Elsevier Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据