期刊
RSC ADVANCES
卷 8, 期 36, 页码 20124-20130出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c8ra02177e
关键词
-
资金
- German Federal Ministry of Education and Research (BMBF, EDIT) [033W010]
- Ministry of Science, Research and Culture (MWFK, digilog-Gesundheitscampus Brandenburg) [06-GeCa: H228-05/002/004]
- German Research Foundation (DFG)
- University of Freiburg
To improve detection sensitivity, molecular diagnostics require preconcentration of low concentrated samples followed by rapid nucleic acid extraction. This is usually achieved by multiple centrifugation, lysis and purification steps, for instance, using chemical reagents, spin columns or magnetic beads. These require extensive infrastructure as well as time consuming manual handling steps and are thus not suitable for point of care testing (POCT). To overcome these challenges, we developed a microfluidic chip combining free-flow electrophoretic (FFE) preconcentration (1 ml down to 5 l) and thermoelectric lysis of bacteria as well as purification of nucleic acids by gel-electrophoresis. The integration of these techniques in a single chip is unique and enables fast, easy and space-saving sample pretreatment without the need for laboratory facilities, making it ideal for the integration into small POCT devices. A preconcentration efficiency of nearly 100% and a lysis/gel-electrophoresis efficiency of about 65% were achieved for the detection of E. coli. The genetic material was analyzed by RT-qPCR targeting the superfolder Green Fluorescent Protein (sfGFP) transcripts to quantify mRNA recovery and qPCR to determine DNA background.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐