3.8 Article

Modified Technique for the Implantation of Berlin Heart Excor Ventricular Assist Device in Children

出版社

SAGE PUBLICATIONS INC
DOI: 10.1177/2150135112437796

关键词

circulatory assist devices; congenital heart surgery; CPB components and configurations; transplantation heart

向作者/读者索取更多资源

The Berlin Heart Excor (BHE) assist device has become our standard mechanical support device for long-term support in children with heart failure. We report two useful surgical modifications for the implantation of the BHE in the pediatric population, improving ease of implantation as well as subsequent surgical procedures at the time of explantation or transplantation. The first modification entails the use of a polytetrafluoroethylene graft for cannulation via the innominate artery for arterial perfusion during cardiopulmonary bypass. This graft is preserved and reused for institution of bypass at the time of transplantation or explantation of the device after recovery. The second modification consists of an extension of the BHE arterial cannula using a length of knitted polyester graft. This allows for improved positioning of the arterial cannula in the ascending aorta or main pulmonary artery, facilitates vascular anastomosis, and improves hemostasis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据