Development of a Pneumatic Tensioning Device for Gap Measurement during Total Knee Arthroplasty

期刊

CLINICS IN ORTHOPEDIC SURGERY
卷 4, 期 3, 页码 188-192

出版社

KOREAN ORTHOPAEDIC ASSOC
DOI: 10.4055/cios.2012.4.3.188

关键词

Total knee arthroplasty; Soft tissue balancing; Gap measurement; Pneumatic tensor

资金

  1. Catholic Medical Center Research Foundation

向作者/读者索取更多资源

Background: Despite the importance of soft tissue balancing during total knee arthroplasty (TKA), all estimating techniques are dependent on a surgeon's manual distraction force or subjective feeling based on experience. We developed a new device for dynamic gap balancing, which can offer constant load to the gap between the femur and tibia, using pneumatic pressure during range of motion. Methods: To determine the amount of distraction force for the new device, 3 experienced surgeons' manual distraction force was measured using a conventional spreader. A new device called the consistent load pneumatic tensor was developed on the basis of the biomechanical tests. Reliability testing for the new device was performed using 5 cadaveric knees by the same surgeons. Intraclass correlation coefficients (ICCs) were calculated. Results: The distraction force applied to the new pneumatic tensioning device was determined to be 150 N. The interobserver reliability was very good for the newly tested spreader device with ICCs between 0.828 and 0.881. Conclusions: The new pneumatic tensioning device can enable us to properly evaluate the soft tissue balance throughout the range of motion during TKA with acceptable reproducibility.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据