期刊
IEEE PHOTONICS JOURNAL
卷 10, 期 1, 页码 -出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JPHOT.2018.2792447
关键词
Self-mixing interferometer; diffraction grating; accelerometer testing
资金
- National Natural Science Foundation of China [51405240]
- Natural Science Research of Jiangsu Higher Education Institutions of China [17KJA510002]
A simple and robust testing system based on a laser self-mixing grating interferometer (SMGI) is proposed to determine the sensitivity of accelerometers. Self-mixing grating interference occurs when the light emitted from a laser diode is incident onto a reflective grating at a fixed angle and the first-order diffracted light returns back into the laser cavity. Frequency ratio method and minimum point method that are officially used to calibrate accelerometers are modified to make it suitable to an SMGI testing system. In order to evaluate the performance of the proposed method, the sensitivity of a commercial microelectromechanical system accelerometer was tested at different vibration frequency. The obtained results show good agreement with the sensitivity given by the manufacturer's specification. Given its stability, simplicity, and efficiency, the proposed system has the potential to be adopted as an alternative method to test accelerometers for industrial applications.