IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS
卷 9, 期 3, 页码 370-376出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TBCAS.2014.2346761
关键词
Cardiovascular monitoring; electrocardiogram; heart rate; motion artifacts; wearable sensor
资金
- MIT Medical Electronic Device Realization Center (MEDRC)
An ASIC for wearable cardiovascular monitoring is implemented using a topology that takes advantage of the electrocardiogram's (ECG) waveform to replace the traditional ECG instrumentation amplifier, ADC, and signal processor with a single chip solution. The ASIC can extract heartbeat timings in the presence of baseline drift, muscle artifact, and signal clipping. The circuit can operate with ECGs ranging from the chest location to remote locations where the ECG magnitude is as low as 30 mu V. Besides heartbeat detection, a midpoint estimation method can accurately extract the ECG R-wave timing, enabling the calculations of heart rate variability. With 58 nW of power consumption at 0.8 mu V supply voltage and 0.76 mm(2) of active die area in standard 0.18 m CMOS technology, the ECG ASIC is sufficiently low power and compact to be suitable for long term and wearable cardiovascular monitoring applications under stringent battery and size constraints.