4.4 Article

Study on binding phenomenon of lipase enzyme with tributyrin on the surface of graphene oxide array using surface plasmon resonance

期刊

THIN SOLID FILMS
卷 645, 期 -, 页码 10-18

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.tsf.2017.10.021

关键词

Self-assembly; Surface plasmon resonance; Graphene oxide; 4-Amino phenol thio phenol; Non-specific interaction; Triglyceride biosensor

向作者/读者索取更多资源

The performance of biological sensors can be significantly improved by understanding the interaction phenomena at the sensing interfaces. Graphene oxide (GO) surface, has oxygen containing functional groups that strongly interact with the enzyme as well as affect the binding phenomenon. This paper reports an SPR-based approach for the development of lipase (LIP) enzyme based triglyceride (TG) biosensor using the graphene oxide (GO) platform, covalently linked on self-assembled monolayer. The GO flakes, prepared by Hummer's modified process was covalently bonded with self-assembled monolayer of 4-amino thiophenol (4-ATP) on gold (Au) to fabricate GO/4-ATP/Au SPR surface. The LIP was covalently immobilized with the carboxylic acid group of GO sheet (LIP/GO/4-ATP/Au) to construct sensing platform (LIP/GO/4-ATP/Au) for the detection of TG using SPR. The SPR angle is found to vary linearly with the concentration of tributyrin (tbn) within a range of 20 mg dL(-1) to 350 mg dL(-1). The value of association equilibrium constant (K-A) is 3.4 mM. This shows that SPR signal is sensitive to detect low concentration of TG i.e. 20 mg dL(-1). The LIP/GO/4-ATP/Au bioelectrode is also fabricated to conduct the cyclic voltammetric (CV) study as supportive information and found to have a linear range of 50 to 350 mg dL(-1) and low K-m value of 0.88 mmol. The SPR surface provides wider linearity, low detection limit than the electrochemical platform and high affinity towards the tbn but suffers from non-specific interaction by the serum substances which provides useful information for commercialization of LIP and GO nano sheet combination based TG biosensors.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据