4.7 Article

A QCM humidity sensor based on fullerene/graphene oxide nanocomposites with high quality factor

期刊

SENSORS AND ACTUATORS B-CHEMICAL
卷 266, 期 -, 页码 534-542

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2018.03.143

关键词

Quartz crystal microbalance; Fullerene; Graphene oxide; Resonant sensor; Humidity sensor

资金

  1. National Natural Science Foundation of China [61471305]
  2. Key Project of National Natural Science Foundation of China [61731016]
  3. Science and Technology Support Program of Sichuan Province [2016JZ0028]

向作者/读者索取更多资源

This paper described a quartz crystal microbalance (QCM) humidity sensor based on fullerene (C-60)/graphene oxide (GO) nanocomposites. Humidity sensing properties such as the quality factor, frequency response property, response/recovery time, and humidity hysteresis were measured and quantitatively compared with a GO based QCM humidity sensor. The experiments provided evidences that the GO film was viscous even at low relative humidity, which had a large impact on the measurable result of a GO coated QCM humidity sensor. As a result, the proposed sensor showed better performances at quality factor and dynamic response/recovery behavior than the GO based QCM humidity sensor. Based on the admittance analysis and frequency response analysis of the sensors, it was speculated that the introduction of C-60 molecules into GO reduced the aggregation of the GO sheets and formed some hydrophobic isolation layers between the GO sheets, which could inhibit the water molecules permeation and maintain the mechanical stiffness of the C-60/GO film. (C) 2018 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据