期刊
SENSORS AND ACTUATORS A-PHYSICAL
卷 276, 期 -, 页码 68-75出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.sna.2018.04.022
关键词
Capnometer; Infrared absorption; Microoptical system; Gas sensor; CO2; Sensor; Clinical ventilation
资金
- Federal Ministry of Education and Research (BMBF) [16SV5606]
In the case of clinical ventilation, sensors are needed that analyze the carbon dioxide concentration during the exhalation cycle. For devices capable to individually control and adjust parameters for each patient, the analysis of the breathing conditions including carbon dioxide concentration is of critical importance. We present a miniaturized system that is optimized for a fast passive gas exchange with low dead volumes in the side stream. The presented system which is also suitable for neonates shows high dynamics and no influence on the breathing behavior due to small systems sizes. Cost reduction is achieved through a fully passive gas flow that does not require active elements such as valves and pumps. The cuvette has been designed for optimized optical and fluidic properties for surveillance in a wide range of breathing conditions. (C) 2018 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐