期刊
SENSORS
卷 18, 期 7, 页码 -出版社
MDPI
DOI: 10.3390/s18072038
关键词
label-free biosensing performance; calculation of the limit of detection; limit of quantification; measurement uncertainty
资金
- European Commission [FP7-OCEAN-2013-614057, H2020-NMBP-768641]
- Spanish Ministry Ministerio de Economia y Competitividad [TEC2017-84846-R, RTC-2015-3273-1]
A significant amount of noteworthy articles reviewing different label-free biosensors are being published in the last years. Most of the times, the comparison among the different biosensors is limited by the procedure used of calculating the limit of detection and the measurement uncertainty. This article clarifies and establishes a simple procedure to determine the calibration function and the uncertainty of the concentration measured at any point of the measuring interval of a generic label-free biosensor. The value of the limit of detection arises naturally from this model as the limit at which uncertainty tends when the concentration tends to zero. The need to provide additional information, such as the measurement interval and its linearity, among others, on the analytical systems and biosensor in addition to the detection limit is pointed out. Finally, the model is applied to curves that are typically obtained in immunoassays and a discussion is made on the application validity of the model and its limitations.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据