4.6 Article Proceedings Paper

An Injectable 64 nW ECG Mixed-Signal SoC in 65 nm for Arrhythmia Monitoring

期刊

IEEE JOURNAL OF SOLID-STATE CIRCUITS
卷 50, 期 1, 页码 375-390

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/JSSC.2014.2364036

关键词

Analog front-end (AFE); asynchronous logic; capacitive feedback chopper stabilized instrumental amplifier (CCIA); electrocardiography (ECG); fast Fourier transform (FFT); implantable system; kickback noise; minimum energy point (MEP); mm(3)

资金

  1. National Heart, Lung, and Blood Institute [R01-HL118304]
  2. Leducq Foundation
  3. University of Michigan
  4. National Science Foundation, DARPA/StartNet Terraswarm Research Center

向作者/读者索取更多资源

A syringe-implantable electrocardiography (ECG) monitoring system is proposed. The noise optimization and circuit techniques in the analog front-end (AFE) enable 31 nA current consumption while a minimum energy computation approach in the digital back-end reduces digital energy consumption by 40%. The proposed SoC is fabricated in 65 nm CMOS and consumes 64 nW while successfully detecting atrial fibrillation arrhythmia and storing the irregular waveform in memory in experiments using an ECG simulator, a live sheep, and an isolated sheep heart.