4.5 Article

In-operando photoelectron spectroscopy for batteries: Set-up using pristine thin film cathode and first results on NaxCoO2

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 89, 期 7, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.5022616

关键词

-

资金

  1. DFG [HA 6128/2-1, KO 4093/2-1]
  2. BMBF [03KP801]

向作者/读者索取更多资源

A model all-solid-state battery cell with a thin film Na,CoO2 cathode was assembled under ultra-high vacuum conditions and cycled inside the vacuum chamber, using a dedicated sample holder. We present in-operando x-ray photoelectron spectroscopy measurements of a NaxCoO2 cathode at different charging states. During battery operation, the change in sodium content, the change in cobalt oxidation state, and the evolution of the O1s and VB emissions could be monitored. Comparison with a conventional post-mortem analysis technique showed that the new measurement technique produces comparable results regarding the oxidation state of the transition metal, but sodium and oxygen results show differences due to cathode/electrolyte interfacial reactions for conventional analysis. By using surface layer-free samples in the presented techniques, we could circumvent such reactions and obtain reliable spectra for the pure bulk-like active cathode material. Published by AIP Publishing.