4.4 Article

A Monte Carlo simulation study of the impact of novel scintillation crystals on performance characteristics of PET scanners

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2018.05.010

关键词

PET; Scintillation crystals; Monte Carlo simulations; NEMA standard; Modelling

资金

  1. Swiss National Science Foundation [SNSF 320030_176052]
  2. Swiss Cancer Research Foundation [KFS-3855-02-2016]

向作者/读者索取更多资源

Objective: The purpose of this study is to validate a Monte Carlo simulation model for the clinical Siemens Biograph mCT PET scanner using the GATE simulation toolkit, and to evaluate the performance of six different scintillation materials in this model using the National Electrical Manufactures Association (NEMA) NU 2-2007 protocol. Methods: A model of the Biograph mCT PET detection system and its geometry was developed. NEMA NU 2-2007 phantoms were also modelled. The accuracy of the developed scanner model was validated through a comparison of the simulation results from GATE, SimSET and PeneloPET toolkits, and experimental data obtained using the NEMA NU 2-2007 protocols. The evaluated performance metrics included count rate performance, spatial resolution, sensitivity, and scatter fraction (SF). Thereafter, the mCT PET scanner was simulated with six different candidate high-performance scintillation materials, including LSO, LaBr3, CeBr3, LuAP, GLuGAG and LFS-3, and its performance evaluated according to the NEMA NU 2-2007 specifications. Results: The Monte Carlo simulation model demonstrates good agreement with the experimental data and results from other simulation packages. For instance, the scatter fraction calculated using GATE simulation is 34.35% while the experimentally measured value is 33.2%, 38.48% for the SimSET, and 34.8% for the PeneloPET toolkit. The best-performing scintillation materials were found to be LuAP, LSO and LFS-3, while GLuGAG offers acceptable performance if cost is the dominant concern. Conclusion: The main performance characteristics of the Biograph mCT PET scanner can be simulated accurately using GATE with a good agreement with other Monte Carlo simulation packages and experimental measurements. Newly developed scintillators show promise and offer alternative options for the design of novel generation PET scanners.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据