4.4 Review

Monte Carlo simulation of eye lens dose reduction from CT scan using organ based tube current modulation

期刊

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2018.03.011

关键词

Monte Carlo; Eye lens dose; Tube current modulation; Eye model

资金

  1. Natural Science Foundation of China [11475047]

向作者/读者索取更多资源

Purpose: To investigate lens dose reduction with organ based tube current modulation (TCM) using the Monte Carlo method. Methods: To calculate lens dose with organ based TCM, 36 pairs of X-ray sources with bowtie filters were placed around the patient head using a projection angle interval of 10 degrees for one rotation of Computed Tomography (CT). Each projection was simulated respectively. Both voxelized and stylized eye models and Chinese reference male phantoms were used in the simulation, and tube voltages 80, 100, 120 and 140 kVp were used. Results: Dose differences between two eye models were less than 20%, but large variations were observed among dose results from different projections of all tube voltages investigated. Dose results from 0 degrees (AP) directions were 60 times greater than those from 180 degrees (PA) directions, which enables organ based TCM reduce lens doses by more than 47%. Conclusions: Organ based TCM may be used to reduce lens doses. Stylized eye models are more anatomically realistic compared with voxelized eye models and are more reliable for dose evaluation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据