期刊
NEW JOURNAL OF CHEMISTRY
卷 42, 期 14, 页码 11944-11953出版社
ROYAL SOC CHEMISTRY
DOI: 10.1039/c7nj04327a
关键词
-
资金
- National Natural Science Foundation of China [21277098]
- Fundamental Research Funds for the Central Universities [1380219166]
An amperometric glucose sensor, PTFE/GOx/AuNPs/PANI/MWCNTs/GCE, was assembled layer-by-layer using multi-wall carbon nanotubes (MWCNTs), polyaniline (PANI), and gold nanoparticles (AuNPs) for electrode surface modification and a fixed glucose oxidase enzyme (GOx) carrier. Methods such as cyclic voltammetry (CV) and chronoamperometry (CA) were used to investigate the performance of the modified materials and the sensor itself. The sensor responded well to glucose, and the enzyme fixed on the sensor experienced direct electron transfer (DET). The redox peak current for GOx exhibited a good linear relationship to the scan rate. The apparent heterogeneous electron transfer rate constant, K, was 5.19 s(-1) (n = 2); the sensor reached 95.0% of the maximum stable current in 10 s. Moreover, the linear range for the glucose concentration was 0.0625 mM to 1.19 mM with a sensitivity of 29.17 A mM(-1) cm(-2) and a detection limit of 0.19 M. The sensor retained 92.0% of its initial activity after 30 days. The average relative standard deviation for the serum sample measurements was 2.3% (n = 3). The recovery rates ranged from 97.3% to 104.1% for various glucose concentrations. The results indicate that the PTFE/GOx/AuNPs/PANI/MWCNTs/GCE sensor can measure the glucose concentration in human serum and holds potential for future clinical applications.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据