4.6 Article

Graphene onto medical grade titanium: an atom-thick multimodal coating that promotes osteoblast maturation and inhibits biofilm formation from distinct species

期刊

NANOTOXICOLOGY
卷 12, 期 4, 页码 274-289

出版社

TAYLOR & FRANCIS LTD
DOI: 10.1080/17435390.2018.1434911

关键词

Carbon; osteogenesis; differentiation; infection; osseointegration; Raman

向作者/读者索取更多资源

The time needed for the osseointegration of titanium implants is deemed too long. Moreover, the bacterial colonization of their surfaces is a major cause of failure. Graphene can overcome these issues but its wet transfer onto substrates employs hazardous chemicals limiting the clinical applications. Alternatively, dry transfer technique has been developed, but the biological properties of this technique remain unexplored. Here, a dry transfer technique based on a hot-pressing method allowed to coat titanium substrates with high-quality graphene and coverage area >90% with a single transfer. The graphene-coated titanium is cytocompatible, did not induce cell membrane damage, induced human osteoblast maturation (gene and protein level), and increased the deposition of mineralized matrix compared to titanium alone. Moreover, graphene decreased the formation of biofilms from Streptococcus mutans, Enterococcus faecalis and even from whole saliva on titanium without killing the bacteria. These findings confirm that coating of titanium with graphene via a dry transfer technique is a promising strategy to improve osseointegration and prevent biofilm formation on implants and devices.