3.8 Article

Pacemaker lead fracture without an increase in lead impedance caused by cardiac fibroma

期刊

JOURNAL OF ARRHYTHMIA
卷 29, 期 6, 页码 357-359

出版社

WILEY
DOI: 10.1016/j.joa.2013.04.001

关键词

Pacemaker lead fracture; Pacemaker lead impedance; Tricuspid valve; Cardiac fibroma

向作者/读者索取更多资源

We report the case of a 64-year-old man who had a permanent pacemaker with a unipolar silicone electrode positioned in the right ventricle in 1989 for sinus node dysfunction. On a routine checkup in June 2011, a 28-mm-diameter mass was discovered, which appeared to adhere to the tricuspid valve and the ventricular lead. The size of the mass did not change for the next 6 months, and the lead impedance was maintained at around 500-600 0. Because pacing failure was observed in January 2012, he underwent an urgent pacemaker check; however, the lead impedance was found not to have increased greatly (689 0). Nevertheless, the pacemaker lead was noted to be fractured at the tricuspid level. His echocardiogram showed new severe tricuspid regurgitation and a floating mass around the lead. We extracted the fractured lead, enucleated the tumor, replaced the tricuspid valve, and placed an epicardial lead. Macroscopic examination revealed that the tumor surrounded the fractured lead and covered the stump. Pathological examination revealed that the tumor was composed of Fibrous connective tissue. We presumed that electric current continued to flow through the stump of the fractured unipolar lead to the generator, and this might have caused the limited increase in lead impedance. (C) 2013 Japanese Heart Rhythm Society. Published by Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据