3.8 Proceedings Paper

Nanoelectrode Chrome Photomask Design and Specification for Biosensor Fabrication

出版社

TRANS TECH PUBLICATIONS LTD
DOI: 10.4028/www.scientific.net/AMR.795.397

关键词

Nanoelectrode; Chrome photomask; Photolithography; AutoCAD; Critical Dimension; Biosensor; Nano Diagnostic; Clinical Diagnostics

向作者/读者索取更多资源

This paper explains the most crucial part of any microchip fabrication, which is the mask design for photolithography process. The design is initially sketched roughly to meet the design specification and later on designed using AutoCAD software. Therefore, to meet the required criteria, the overall width and length of the device is optimized at 12mm and 20.21mm respectively. Optimization of the size is done based on the chip behavior as a disposable chip and adding an economical value when it is commercialized. The nano electrode mask layout comprises of four sets of design which are single gaps for size reduction, single gaps for size expansion, multiple gaps for size reduction and multiple gaps for size expansion. While, the second chrome mask is fabricated for gold contact padding with two types of design sets, one is for single gaps and another is for multiple gaps. Both mask designs were sent for chrome mask fabrication for future use in biosensor fabrication.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据