4.5 Article

A Sortase A-Immobilized Mesoporous Hollow Carbon Sphere-Based Biosensor for Detection of Gram-Positive Bacteria

期刊

JOURNAL OF ELECTRONIC MATERIALS
卷 47, 期 7, 页码 4124-4135

出版社

SPRINGER
DOI: 10.1007/s11664-018-6308-4

关键词

Staphylococcus aureus; sortase A; immobilization; mesoporous carbon material; biosensor

资金

  1. National Natural Science Foundation of China [31572566, 31402251]

向作者/读者索取更多资源

A sensor based on mesoporous carbon materials immobilized with sortase A (SrtA) for determination of Staphylococcus aureus (S. aureus) is reported. To prepare the biosensor, we first synthesized carboxyl-functionalized mesoporous hollow carbon spheres, then applied them as carriers for immobilization of SrtA. Based on the catalytic mechanism of SrtA, a highly sensitive, inexpensive, and rapid method was developed for S. aureus detection. The sensor showed a linear response in the bacterial concentration range of 0.125 x 10(2) colony-forming units (CFU) mL(-1) to 2.5 x 10(2) CFU mL(-1), with detection limit as low as 9.0 CFU mL(-1). The method was successfully used for quantitative detection of S. aureus in whole milk samples, giving results similar to experimental results obtained from the plate counting method. This biosensor could also be used to detect other Gram-positive bacteria that secrete SrtA.