Investigating optical path and differential pathlength factor in reflectance photoplethysmography for the assessment of perfusion

期刊

JOURNAL OF BIOMEDICAL OPTICS
卷 23, 期 7, 页码 -

出版社

SPIE-SOC PHOTO-OPTICAL INSTRUMENTATION ENGINEERS
DOI: 10.1117/1.JBO.23.7.075005

关键词

photoplethysmography; Beer-Lambert law; Monte Carlo model; optical path; near-infrared spectroscopy; oxy/deoxy hemoglobin

向作者/读者索取更多资源

Photoplethysmography (PPG) is an optical noninvasive technique with the potential for assessing tissue perfusion. The relative time-change in the concentration of oxyhemoglobin and deoxyhemoglobin in the blood can be derived from DC part of the PPG signal. However, the absolute concentration cannot be determined due to the inadequate data on PPG optical paths. The optical path and differential pathlength factor (DPF) for PPG at red (660 nm) and infrared (880 nm) wavelengths were investigated using a heterogeneous Monte Carlo model of the human forearm. Using the simulated DPFs, the absolute time-change in concentrations were determined from PPG signals recorded from the same tissue site. Results were compared with three conditions of approximated DPFs. Results showed the variation of the optical-path and DPF with different wave-lengths and source-detector separations. Approximations resulted in significant errors, for example, using NIRS DPF in PPG led to cross talk of -0.4297 and 0.060 and an error of 15.16% to 25.18%. Results confirmed the feasibility of using the PPG (DC) for the assessment of tissue perfusion. The study also identified the inappropriateness of the assumption that DPF is independent of wavelength or source-detector separations and set the platform for further studies on investigating optical pathlengths and DPF in PPG. (C) 2018 Society of Photo-Optical Instrumentation Engineers (SPIE).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据