4.5 Article

Original article submission: Platelet stress accumulation analysis to predict thrombogenicity of an artificial kidney

期刊

JOURNAL OF BIOMECHANICS
卷 69, 期 -, 页码 26-33

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.jbiomech.2018.01.014

关键词

Thrombogenicity; Artificial kidney; Platelet stress accumulation; Hemodynamics

资金

  1. National Institutes of Health [1R01EB014315, 1U01EB012487, 1U01EB021214]

向作者/读者索取更多资源

An implantable artificial kidney using a hemofilter constructed from an array of silicon membranes to provide ultrafiltration requires a suitable blood flow path to ensure stable operation in vivo. Two types of flow paths distributing blood to the array of membranes were evaluated: parallel and serpentine. Computational fluid dynamics (CFD) simulations were used to guide the development of the blood flow paths. Pressure data from animal tests were used to obtain pulsatile flow conditions imposed in the transient simulations. A key consideration for stable operation in vivo is limiting platelet stress accumulation to avoid platelet activation and thrombus formation. Platelet stress exposure was evaluated by CFD particle tracking methods through the devices to provide distributions of platelet stress accumulation. The distributions of stress accumulation over the duration of a platelet lifetime for each device revealed that stress accumulation for the serpentine flow path exceeded levels expected to cause platelet activation while the accumulated stress for the parallel flow path was below expected activation levels. (C) 2018 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据