A peak cut procedure of column separation for calcium isotope measurement using the double spike technique and thermal ionization mass spectrometry (TIMS)

期刊

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
卷 33, 期 4, 页码 547-554

出版社

ROYAL SOC CHEMISTRY
DOI: 10.1039/c8ja00004b

关键词

-

资金

  1. National Natural Science Foundation of China [41490632, 41773009, 41773062]
  2. National Key R&D Program of China [2016YFC0600408]

向作者/读者索取更多资源

Full recovery from column separation and matrix effects are the two factors that need to be considered for the high-precision analysis of stable Ca isotopes, but generally they are difficult to balance. In many cases, to get a pure Ca fraction, the interference of the matrix elements is reduced at the cost of discarding a fraction of Ca overlapping with other elements (e.g. Sr and K). However, quantitative evaluation using this approach is challenging but greatly needed. Our study investigates the influence of low Ca recovery on delta Ca-44/40 using thermal ionization mass spectrometry (TIMS) with the double spike technique. delta Ca-44/40 of IAPSO seawater, ML3B-G and BHVO-2 in different Ca subcuts (e.g. 0-20, 20-40, 40-60, 6080 and 80-100%), display limited variation after iterative correction by Ca-42-Ca-43 double spike with the exponential law. Notably, delta Ca-44/40 of each Ca subcut with similar to 20% recovery is consistent with that of the Ca cut with full recovery, within a margin of error. Our results indicate that Ca-42-Ca-43 double spike technique can simultaneously correct Ca isotopic fractionation, occurring during column separation, and TIMS determination, because both follow the exponential fractionation law well. Therefore, a peak cut procedure of column separation for Ca isotope measurement using the double spike technique on TIMS is proposed. Briefly, we can mix the double spike with the sample solution well before column separation, then collect the peak of the Ca cut and abandon both sides of the Ca eluate that may overlap with other elements. This procedure would eliminate matrix effects efficiently, especially for samples with low CaO contents which typically must be passed through the column twice (e.g. peridotite and dunite).

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据