期刊
出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/j.proche.2014.10.041
关键词
HPLC; lactide; glycolide; biodegradable polymers
A simple, rapid high-performance liquid chromatographic (HPLC) method was developed for the simultaneous determination of glycolic acid, lactic acid, glicolide, lactide and ethyacetate in monomers for obtaining biopolymers. The separation was effected on the reversed-phase C18 column 250mmx4.6mm with particle size 5 mu using a mobile phase mixture buffer and acetonitrile in a ratio 88: 12 v/v and elution was isocratic at a flow-rate of 1.0 mL/min. The determinations were performed with a UV-Vis detector at 200 nm. The volume of the injected sample was 20 mu L. Detection limits for acids and its dimers (glycolic acid, lactic acid, glicolide, lactide) and ethylacetate range between 82 and 182 ng/mL. The analytes are separated in 13 min. Recovery studies showed good results for all solutes (99-102%). The method is linear for all compounds over the concentration range tested, and shows good precision and accuracy, making it suitable for quantitation of acids and its dimers (glycolic acid, lactic acid, glicolide, lactide) and ethyl acetate in monomers. (C) 2014 The Authors. Published by Elsevier B.V.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐