期刊
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT
卷 67, 期 2, 页码 431-438出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TIM.2017.2779318
关键词
Electronic tongue (ET); field-programmable gate array (FPGA); liquid sensor; quartz-crystal microbalance (QCM); sensor array
In this paper, an electronic tongue (ET) based on a quartz-crystal microbalance (QCM) sensor array was designed and the methods for sensor data analysis were described. In this regard, as chemical sensors, QCM crystals coated with three different phthalocyanines were employed to discriminate toxic chemicals in liquid media. For sensors' preparation, an automatic coating instrument was developed based on the airbrush technique. The oscillator circuits used in the coating instrument and measurement setup were optimized to minimize the noise and mutual interference of the other oscillator circuits in the sensor array. The designed measurement system, which was responsible for collecting and storing data, communicating with a computer, consisted of a field-programmable gate-array-based control unit. The measurement system collected training data from the sensor array, and then computer application using MATLAB used an artificial neural network to find the composition of toxic chemicals in the sample water. The measurement system was tested for three different toxic chemicals, and the average success rate in predicting them was 4.08%. The desired chemicals can be detectable by preparing the necessary sensor array with the proposed ET.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据