期刊
IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS
卷 66, 期 5, 页码 4108-4117出版社
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TIE.2018.2840522
关键词
Cardiac pacemaker; micro medical robotics; wireless charging; wireless power transfer (WPT)
资金
- ITF Tier3 of Innovation and Technology Commission of HKSAR, China [ITS/353/16]
- Energy and Conservation Fund of Hong Kong Special Administrative Region, China (ECF Project) [92/2016]
This paper presents a new effective sandwiched wireless power transfer (WPT) system, which is adopted to recharge the battery of a micro medical robotics for a cardiac pacemaker. The key of the design is to use the distinct sandwiched topology in both the transmitter and receiver coils, whose operating frequency of 160 kHz is significantly lower than the reported WPT system up to megahertz. Furthermore, the implantable receiver adopts the bilateral coil design, which can harness more power within a limited size. Meanwhile, with the sandwiched transmitting coils, the proposed design achieves the high feasibility and flexibility of the controllable distance for different kinds of micro medical robotics in patients. As a result, the proposed system can provide the power as high as 5 W and the transmission efficiency up to 88%. Finally, the corresponding theoretical analysis, simulation, and experimentation are carried out to prove the validity of the proposed sandwiched medical WPT system for the implantable device in the patient body.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据