Validity of Automated Choroidal Segmentation in SS-OCT and SD-OCT

期刊

INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE
卷 56, 期 5, 页码 3202-3211

出版社

ASSOC RESEARCH VISION OPHTHALMOLOGY INC
DOI: 10.1167/iovs.14-15669

关键词

choroid; automated segmentation; quantification; swept-source OCT; spectral-domain OCT

资金

  1. Netherlands Organization for Scientific Research, the Hague
  2. Swart van Essen, Rotterdam
  3. Bevordering van Volkskracht, Rotterdam
  4. Rotterdamse Blindenbelangen Association, Rotterdam
  5. Algemene Nederlandse Vereniging ter Voorkoming van Blindheid, Doorn, The Netherlands
  6. Oogfonds Nederland, Utrecht
  7. MDFonds, Utrecht
  8. Vereniging Trustfonds Erasmus Universiteit Rotterdam, Rotterdam, The Netherlands
  9. Lijf en Leven, Krimpen aan de IJssel, The Netherlands
  10. Topcon Europe BV, Capelle aan den IJssel, The Netherlands
  11. [R01-EY017066]
  12. [R01-EY018853]
  13. [R01-EB004640]

向作者/读者索取更多资源

PURPOSE. To evaluate the validity of a novel fully automated three-dimensional (3D) method capable of segmenting the choroid from two different optical coherence tomography scanners: swept-source OCT (SS-OCT) and spectral-domain OCT (SD-OCT). METHODS. One hundred eight subjects were imaged using SS-OCT and SD-OCT. A 3D method was used to segment the choroid and quantify the choroidal thickness along each A-scan. The segmented choroidal posterior boundary was evaluated by comparing to manual segmentation. Differences were assessed to test the agreement between segmentation results of the same subject. Choroidal thickness was defined as the Euclidian distance between Bruch's membrane and the choroidal posterior boundary, and reproducibility was analyzed using automatically and manually determined choroidal thicknesses. RESULTS. For SS-OCT, the average choroidal thickness of the entire 6- by 6-mm(2) macular region was 219.5 mu m (95% confidence interval [CI], 204.9-234.2 mu m), and for SD-OCT it was 209.5 mu m (95% CI, 197.9-221.0 mu m). The agreement between automated and manual segmentations was high: Average relative difference was less than 5 lm, and average absolute difference was less than 15 lm. Reproducibility of choroidal thickness between repeated SS-OCT scans was high (coefficient of variation [CV] of 3.3%, intraclass correlation coefficient [ICC] of 0.98), and differences between SS-OCT and SD-OCT results were small (CV of 11.0%, ICC of 0.73). CONCLUSIONS. We have developed a fully automated 3D method for segmenting the choroid and quantifying choroidal thickness along each A-scan. The method yielded high validity. Our method can be used reliably to study local choroidal changes and may improve the diagnosis and management of patients with ocular diseases in which the choroid is affected.