4.4 Article

Accuracy of a Factory-Calibrated, Real-Time Continuous Glucose Monitoring System During 10 Days of Use in Youth and Adults with Diabetes

期刊

DIABETES TECHNOLOGY & THERAPEUTICS
卷 20, 期 6, 页码 395-402

出版社

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2018.0150

关键词

Glucose sensor performance; Clinical accuracy; MARD; Continuous glucose monitoring; Advanced algorithm; Factory-calibrated

资金

  1. Dexcom, Inc. (San Diego, CA)

向作者/读者索取更多资源

Background: Frequent use of continuous glucose monitoring (CGM) systems is associated with improved glycemic outcomes in persons with diabetes, but the need for calibrations and sensor insertions are often barriers to adoption. In this study, we evaluated the performance of G6, a sixth-generation, factory-calibrated CGM system specified for 10-day wear. Methods: The study enrolled participants of ages 6 years and up with type 1 diabetes or insulin-treated type 2 diabetes at 11 sites in the United States. Participation involved one sensor wear period of up to 10 days. Adults wore the system on the abdomen; youth of ages 6-17 years could choose to wear it on the abdomen or upper buttocks. Clinic sessions for frequent comparison with reference blood glucose measurements took place on days 1, 4-5, 7, and/or 10. Participants of ages 13 years and up underwent purposeful supervised glucose manipulation during in-clinic sessions. During the study, participants calibrated the systems once daily. However, analysis was performed on glucose values that were derived from reprocessed raw sensor data, independently of self-monitored blood glucose values used for calibration. Reprocessing used assigned sensor codes and a factory-calibration algorithm. Performance evaluation included the proportion of CGM values that were within 20% of reference glucose values >100mg/dL or within 20mg/dL of reference glucose values 100mg/dL (%20/20), the analogous %15/15, and the mean absolute relative difference (MARD, expressed as a percentage) between temporally matched CGM and reference values. Results: Data from 262 study participants (21,569 matched CGM reference pairs) were analyzed. The overall %15/15, %20/20, and MARD were 82.4%, 92.3%, and 10.0%, respectively. Matched pairs from 134 adults and 128 youth of ages 6-17 years were similar with respect to %20/20 (92.4% and 91.9%) and MARD (9.9% and 10.1%). Overall %20/20 values on days 1 and 10 of sensor wear were 88.6% and 90.6%, respectively. The system's Urgent Low Soon (predictive of hypoglycemia within 20min) hypoglycemia alert was correctly provided 84% of the time within 30min before impending biochemical hypoglycemia (<70mg/dL). The 10-day sensor survival rate was 87%. Conclusion: The new factory-calibrated G6 real-time CGM system provides accurate readings for 10 days and removes several clinical barriers to broader CGM adoption.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据