3.8 Article

Percutaneous Mechanical Ventricular Support in Acute Cardiac Care: A UK Quaternary Centre Experience Using 2.5L, 3.8L and 5.0L Impella Catheters

期刊

CARDIOLOGY AND THERAPY
卷 4, 期 1, 页码 47-58

出版社

SPRINGER LONDON LTD
DOI: 10.1007/s40119-014-0033-8

关键词

High-risk PCI; Impella 3.8L; Impella device; Outcomes; Percutaneous coronary intervention; Vascular complications

向作者/读者索取更多资源

Aims: The Impella is a percutaneous ventricular assist device. The majority of published data describes the 2.5L and 5.0L devices, and little data is available for the newer 3.8L device. We examined the indications and outcomes from our single-centre real-world'' registry at The Queen Elizabeth Hospital, Birmingham, UK, using all three pump sizes. Methods and Results: Records from all patients who underwent attempted Impella-assisted procedures at our centre were examined retrospectively. Impella implantation was attempted in 49 patients (mean age 72 +/- 13 years; 80% male) and was successful in 48 (98%). 45 patients underwent high-risk percutaneous coronary intervention (PCI), one patient underwent balloon aortic valvuloplasty and 3 patients had Impella as a bridge to cardiac transplantation. The 2.5L and 3.8L devices were used in 36 (75%) and 11 (23%) patients, respectively, while one patient (2%) had the 5L device. Vascular complications occurred in only one patient (2%) and stroke and peri-procedural myocardial infarction occurred in one patient (2%), while in-hospital mortality was 20% (10/49). Conclusions: In this large real-world registry, we have demonstrated the safety and feasibility of the Impella device for a wide range of indications. This includes the first series of the 3.8L device which provides superior support with no increase in vascular complications.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据