4.7 Article

Application of handheld and portable spectrometers for screening acrylamide content in commercial potato chips

期刊

FOOD CHEMISTRY
卷 174, 期 -, 页码 154-162

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.foodchem.2014.11.001

关键词

Acrylamide; Potato chips; Handheld spectrometers; Portable spectrometers; NIR; MIR

资金

  1. Turkish Ministry of Education
  2. Ohio Agricultural Research and Development Center

向作者/读者索取更多资源

The most common methods for acrylamide analysis in foods require the use of LC-MS/MS and GC-MS. Although these methods have great analytical performance, they need intensive sample preparation, highly specialised instrumentation, and are time consuming. In this study, portable and handheld infrared spectrometers were evaluated as rapid methods for screening acrylamide in potato chips and their performances were compared to those of benchtop infrared systems. The acrylamide content of 64 commercial potato chips (169-2453 mu g/kg) was determined by LC-MS/MS. Spectral data were collected using mid-infrared (MIR) and near-infrared (NIR) spectrometers. Partial least squares regression (PLSR) calibration models were developed to predict acrylamide levels. Overall, good linear correlation was found between the predicted acrylamide levels and actual measured acrylamide concentrations by LC-MS/ MS (rPred > 0.90 and SEP < 100 mu g/kg). Our results indicate that portable and handheld spectrometers can be used as simple and rapid alternatives for acrylamide analysis in potato chips. (C) 2014 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据