期刊
ORIENTAL JOURNAL OF CHEMISTRY
卷 31, 期 -, 页码 201-206出版社
ORIENTAL SCIENTIFIC PUBL CO
DOI: 10.13005/ojc/31.Special-Issue1.24
关键词
Cyclic voltammetry; paracetamol; phosphate buffer
A sensitive electroanalytical method for determination of acetaminophen (ACT) in paracetamol tablet has been developed and validated. Cyclic voltammetry (CV) was used to study the electrochemical behaviour of paracetamol, while differential pulse voltammetry (DPV) was used to determine ACT. Screen printed carbon electrode was used for the analysis. Determination of all parameters in paracetamol tablet was performed in phosphate buffer (PBS) from pH 4.0 to 9.0. The effects of various factors, such as the scan rates, pH were discussed. For assay of ACT, method validation steps such as linearity, precision, accuracy, limit of detection, limit of quantification were performed.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。