4.8 Article

A fully automated microfluidic-based electrochemical sensor for real-time bacteria detection

期刊

BIOSENSORS & BIOELECTRONICS
卷 100, 期 -, 页码 541-548

出版社

ELSEVIER ADVANCED TECHNOLOGY
DOI: 10.1016/j.bios.2017.09.046

关键词

Real-time pathogen detection; Escherichia coil; Microfluidic-based electrochemical sensor; Amperometry; Waterborne diseases

资金

  1. TUBITAK-BIDEB [BIDES-2221, 2221]

向作者/读者索取更多资源

A fully automated microfluidic-based electrochemical biosensor was designed and manufactured for pathogen detection. The quantification of Escherichia coil was investigated with standard and nanomaterial amplified immunoassays in the concentration ranges of 0.99 x 10(4) -3.98 x 10(9) cfu mL(-1) and 10-3.97 x 10(7) cfu mL(-1) which resulted in detection limits of 1.99 x 10(4) cfu mL(-1) and 50 cfu mL(-1), respectively. The developed methodology was then applied for E. coil quantification in water samples using nanomaterial modified assay. Same detection limit for E. coil was achieved for real sample analysis with a little decrease on the sensor signal. Cross-reactivity studies were conducted by testing Shigella, Salmonella spp., Salmonella typhimurium and Staphylococcus aureus on E. coli specific antibody surface that confirmed the high specificity of the developed immunoassays. The sensor surface could be regenerated multiple times which significantly reduces the cost of the system. Our custom-designed biosensor is capable of detecting bacteria with high sensitivity and specificity, and can serve as a promising tool for pathogen detection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据