4.5 Article

A Finite Element Method to Predict Adverse Events in Intracranial Stenting Using Microstents: In Vitro Verification and Patient Specific Case Study

期刊

ANNALS OF BIOMEDICAL ENGINEERING
卷 44, 期 2, 页码 442-452

出版社

SPRINGER
DOI: 10.1007/s10439-015-1505-2

关键词

Intra-cranial; Cerebral; Aneurysm; Stenting; Hugging; Apposition; Microstent; Incomplete

资金

  1. Research Foundation-Flanders (FWO Grant) [3G065910]

向作者/读者索取更多资源

Clinical studies have demonstrated the efficacy of stent supported coiling for intra-cranial aneurysm treatment. Despite encouraging outcomes, some matters are yet to be addressed. In particular closed stent designs are influenced by the delivery technique and may suffer from under-expansion, with the typical effect of hugging the inner curvature of the vessel which seems related to adverse events. In this study we propose a novel finite element (FE) environment to study potential failure able to reproduce the microcatheter pull-back delivery technique. We first verified our procedure with published in vitro data and then replicated the intervention on one patient treated with a 4.5 x 22 mm Enterprise microstent (Codman Neurovascular; Raynham MA, USA). Results showed good agreement with the in vitro test, catching both size and location of the malapposed area. A simulation of a 28 mm stent in the same geometry highlighted the impact of the delivery technique, which leads to larger area of malapposition. The patient specific simulation matched the global stent configuration and zones prone to malapposition shown on the clinical images with difference in tortuosity between actual and virtual treatment around 2.3%. We conclude that the presented FE strategy provides an accurate description of the stent mechanics and, after further in vivo validation and optimization, will be a tool to aid clinicians to anticipate the acute procedural outcome avoiding poor initial results.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据