4.4 Article

Design of a nitrogen purification system with cryogenic method for neutrino detection

期刊

APPLIED RADIATION AND ISOTOPES
卷 137, 期 -, 页码 194-198

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.apradiso.2018.04.009

关键词

Purification system; Monte Carlo; Liquid scintillator; Cryogenic; Nitrogen

资金

  1. Ministry of Finance of the Peoples Republic of China [ZDYZ2014-1]

向作者/读者索取更多资源

In order to detect the neutrino with liquid scintillation detector, high-purity nitrogen is essential for gas stripping in this detector. Therefore, it is necessary to design a purification system for the detector to purify nitrogen. Using the method of low temperature adsorption for the purification system, the key designs including the flow path, the adsorber and the selection of activated carbon, are introduced in this study. In these designs, the selection of activated carbon is the most important because the adsorption characteristic of the carbon is related to the performance of the purification system. The method of grand canonical ensemble Monte Carlo is adopted to simulate the adsorption of radon by the activated carbon with its slit pore model. Using this method, the working temperature and the key characteristic of the activated carbon are obtained.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据