期刊
REGENERATIVE BIOMATERIALS
卷 3, 期 3, 页码 149-157出版社
OXFORD UNIV PRESS
DOI: 10.1093/rb/rbv030
关键词
cardiovascular devices; biocompatibility; surface modification; hyaluronan; Type IV collagen
The biocompatibility of cardiovascular devices has always been considered crucial for their clinical efficacy. Therefore, a biofunctional coating composed of Type IV collagen (CoIV) and hyaluronan (HA) was previously fabricated onto the titanium (Ti) substrate for the application of promoting vascular smooth muscle cell contractile phenotype and improving surface endothelialization. However, the anti-inflammation property, blood compatibility and in vivo tissue compatibility of the HA/CoIV coating, as paramount consideration of cardiovascular materials surface coating, have not been investigated. Thus, in this study, the three crucial properties of the HA/CoIV coating were tested. The platelet adhesion/activation test and the dynamic whole blood experiment implied that the HA/CoIV coating had better blood compatibility compared with Ti substrate and pure CoIV coating. The macrophage adhesion/activation and inflammatory cytokine release (tumor necrosis factor-alpha and interleukin-1) results indicated that the HA/CoIV coating could significantly improve the anti-inflammation property of the Ti substrate. The in vivo implantation of SD rats for 3 weeks' results demonstrated that the HA/CoIV coating caused milder tissue response. All these results suggested that the multi-functional HA/CoIV coating possessed good biocompatibility. This research is anticipated to be potentially applied for the surface modification of cardiovascular stents.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐