4.3 Article

Expanding metabolite coverage of real-time breath analysis by coupling a universal secondary electrospray ionization source and high resolution mass spectrometry-a pilot study on tobacco smokers

期刊

JOURNAL OF BREATH RESEARCH
卷 10, 期 1, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/1752-7155/10/1/016010

关键词

breath analysis; ionizer; mass spectrometry; real-time; heavy volatiles; high mass resolution

资金

  1. European Community's Seventh Framework Programme (FP7-IAPP) project 'Analytical Chemistry Instrumentation Development' [609691]

向作者/读者索取更多资源

Online breath analysis is an attractive approach to track exhaled compounds without sample preparation. Current commercially available real-time breath analysis platforms require the purchase of a full mass spectrometer. Here we present an ion source compatible with virtually any preexisting atmospheric pressure ionization mass spectrometer that allows real-time analysis of breath. We illustrate the capabilities of such technological development by upgrading an orbitrap mass spectrometer. As a result, we detected compounds in exhaled breath between 70 and 900 Da, with a mass accuracy of typically < 1 ppm; resolutions between m/Delta m 22 000 and 70000 and fragmentation capabilities. The setup was tested in a pilot study, comparing the breath of smokers (n = 9) and non-smokers (n = 10). Exogenous compounds associated to smoking, as well as endogenous metabolites suggesting increased oxidative stress in smokers, were detected and in some cases identified unambiguously. Most of these compounds correlated significantly with smoking frequency and allowed accurate discrimination of smokers and non-smokers.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据