4.4 Article

Liquid Cathode Glow Discharge as a Microplasma Excitation Source for Atomic Emission Spectrometry for the Determination of Trace Heavy Metals in Ore Samples

期刊

ANALYTICAL LETTERS
卷 51, 期 13, 页码 2128-2140

出版社

TAYLOR & FRANCIS INC
DOI: 10.1080/00032719.2017.1406492

关键词

Atomic emission spectrometry; liquid cathode glow discharge; ore samples; trace heavy metals

向作者/读者索取更多资源

A novel liquid cathode glow discharge (LCGD) was designed as a microplasma excitation source for atomic emission spectrometry (AES) and used for the determination of Pb, Cu, and Cd in digested ore samples. The operating parameters and stability of LCGD-AES were investigated in detail. The statistical analysis (t-test) was used for comparing the results of ore samples using LCGD-AES and inductively coupled plasma atomic emission spectrometry (ICP-AES). The results showed that the optimized analytical conditions are 650V discharge voltage, 4.5mLmin(-1) solution flow rate, and pH 1 HNO3 as the supporting electrolyte. The limits of detection for Pb, Cu, and Cd were 0.019, 0.47, and 0.37mgL(-1), respectively. The correlation coefficients and relative standard deviations were 0.9985 and 1.19% for Pb, 0.9868 and 2.37% for Cu, and 0.9960 and 3.98% for Cd. The power consumption was below 65W. The recoveries were in the range from 79.1 to 115.1%. Moreover, the measurement results of ore samples by LCGD-AES are comparable to the values obtained by ICP-AES. Therefore, the LCGD-AES may be developed as a portable analytical instrument for the direct determination of trace heavy metals.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据