Comparison of saddle-shape flexibility and elliptical-shape stability between Cosgrove-Edwards and Memo-3D annuloplasty rings using three-dimensional analysis software

期刊

GENERAL THORACIC AND CARDIOVASCULAR SURGERY
卷 64, 期 6, 页码 325-332

出版社

SPRINGER JAPAN KK
DOI: 10.1007/s11748-016-0645-0

关键词

Cosgrove-Edwards annuloplasty ring; Memo-3D annuloplasty ring; Three-dimensional transesophageal echocardiography; Mitral annular dynamics; Mitral repair

向作者/读者索取更多资源

Objective To compare three-dimensional dynamics between implanted Cosgrove-Edwards and Sorin Memo-3D annuloplasty rings during the cardiac cycle. Methods We examined 11 Cosgrove-Edwards rings and 20 Sorin Memo-3D rings after mitral plasty using real-time three-dimensional transesophageal echocardiography. We evaluated ring height, ellipticity, and geometry during one cardiac cycle. Four evenly spaced phases each selected during systole and diastole were assessed using REAL VIEW software. Results The height of the Cosgrove-Edwards and Sorin Memo-3D rings was similar (2.3 +/- 0.8 vs. 1.9 +/- 0.9 mm, p = 0.44). The maximum difference in ring height during one cardiac cycle (change in height) was larger for the Cosgrove-Edwards than the Sorin Memo-3D rings (2.3 +/- 0.8 vs. 1.5 +/- 0.6 mm, p = 0.014). Ellipticity and the maximum difference in ellipticity during one cardiac cycle (change in ellipticity) were larger for Cosgrove-Edwards than Sorin Memo-3D rings (80.0 +/- 9.1 vs. 72.0 +/- 4.8 %, p = 0.014, respectively, and 12.0 +/- 3.1 vs. 6.0 +/- 1.8 %, p < 0.001). Conclusions Cosgrove-Edwards rings were more flexible, whereas Sorin Memo-3D rings maintained the elliptical shape more effectively.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.0
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据