4.6 Article

Effects of Intraocular Lens Opacification on Light Scatter, Stray Light, and Overall Optical Quality/Performance

期刊

INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE
卷 57, 期 7, 页码 3239-3247

出版社

ASSOC RESEARCH VISION OPHTHALMOLOGY INC
DOI: 10.1167/iovs.16-19514

关键词

intraocular lens; straylight; forward light scattering

资金

  1. Research to Prevent Blindness, Inc. (New York, NY, USA)
  2. Alcon Laboratories (Fort Worth, TX, USA)

向作者/读者索取更多资源

PURPOSE. To evaluate light scatter and stray light in intraocular lenses (IOLs) explanted because of postoperative opacification (13 calcified hydrophilic acrylic, 1 calcified silicone, and 4 polymethylmethacrylate [PMMA] lenses with snowflake degeneration), as well as effect of opacification on other optical quality/performance indicators, in comparison with controls. METHODS. The Complete Angle Scatter Instrument (CASI) scatterometer was used to measure the forward light scattering (FLS) of the IOLs, and the stray light values at various angles were calculated from the measured FLS. Modulation transfer function (MTF) was obtained with an optical bench, and a Badal optometer was used to obtain letter chart images through the lenses. Back light scatter and light transmittance were also measured. RESULTS. Average stray light values (Log (s)) at a scattered angle of 100 were 1.79 +/- 0.37 for hydrophilic acrylic IOLs (controls 0.36 +/- 0.05), 1.53 for the silicone lens (control 0.41), and 1.62 +/- 0.46 for PMMA IOLs (control 0.25). Stray light was significantly higher for explanted opacified lenses (N = 18) in comparison with controls (N = 7; two-tail P < 0.001 at 100). Modulation transfer function and Badal image contrast were drastically reduced in lenses with calcification and snowflake degeneration. CONCLUSIONS. Different studies described the impact of stray light in human vision, with serious hindrance above 1.47 Log (s). Lenses explanted from patients because of clinically significant opacification are associated with a considerable increase in light scatter and stray light, as well as with a decline of other optical quality/performance indicators.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据