4.7 Article

A Flow SPR Immunosensor Based on a Sandwich Direct Method

期刊

BIOSENSORS-BASEL
卷 6, 期 2, 页码 -

出版社

MDPI
DOI: 10.3390/bios6020022

关键词

ampicillin; flow immunosensor SPR; sandwich method

资金

  1. Sapienza University of Rome, Protezione dell'Ambiente e dei Beni Culturali (CIABC) center
  2. Istituto per lo studio dei Materiali Nanostrutturati (ISMN) of CNR (Consiglio Nazionale delle Ricerche)

向作者/读者索取更多资源

In this study, we report the development of an SPR (Surface Plasmon Resonance) immunosensor for the detection of ampicillin, operating under flow conditions. SPR sensors based on both direct (with the immobilization of the antibody) and competitive (with the immobilization of the antigen) methods did not allow the detection of ampicillin. Therefore, a sandwich-based sensor was developed which showed a good linear response towards ampicillin between 10(-3) and 10(-1) M, a measurement time of <= 20 min and a high selectivity both towards beta-lactam antibiotics and antibiotics of different classes.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据