4.2 Article

Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls

期刊

出版社

SPRINGER
DOI: 10.1007/s12178-016-9322-2

关键词

Corrosion; Modular neck; Revision hip arthroplasty; Adverse local tissue reaction

向作者/读者索取更多资源

The management of the patient with a recalled, modular neck-body total hip arthroplasty can be complex, as it involves a combination of clinical, technical, and medicolegal challenges. Management begins with a thorough history and physical exam, radiographic evaluation, infection workup, and serum metal ion levels. Three-dimensional imaging is obtained based on patient symptomatology and metal ion levels and is used to evaluate for the presence of an adverse local tissue response as well as the integrity of the existing soft tissue envelope. The decision to perform revision surgery is based on a combination of patient symptomatology, laboratory values, and imaging findings. Revision surgery involves the entire armamentarium of femoral revision techniques, and the acetabulum may need to be revised at the surgeon's discretion. The femoral implant can often be removed without disrupting the femoral bone envelope; however, the surgeon should have a low threshold to perform an extended trochanteric osteotomy.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据