Advancements, measurement uncertainties, and recent comparisons of the NOAA frost point hygrometer

期刊

ATMOSPHERIC MEASUREMENT TECHNIQUES
卷 9, 期 9, 页码 4295-4310

出版社

COPERNICUS GESELLSCHAFT MBH
DOI: 10.5194/amt-9-4295-2016

关键词

-

资金

  1. US GCOS Program
  2. NASA Upper Atmospheric Research Program
  3. EMRP project METEOMET [ENV07]
  4. EMRP participating countries within EURAMET
  5. European Union

向作者/读者索取更多资源

The NOAA frost point hygrometer (FPH) is a balloon-borne instrument flown monthly at three sites to measure water vapor profiles up to 28 km. The FPH record from Boulder, Colorado, is the longest continuous stratospheric water vapor record. The instrument has an uncertainty in the stratosphere that is <6% and up to 12% in the troposphere. A digital microcontroller version of the instrument improved upon the older versions in 2008 with sunlight filtering, better frost control, and resistance to radio frequency interference (RFI). A new thermistor calibration technique was implemented in 2014, decreasing the uncertainty in the thermistor calibration fit to less than 0.01 degrees C over the full range of frost - or dew point temperatures (-93 to +20 degrees C) measured during a profile. Results from multiple water vapor intercomparisons are presented, including the excellent agreement between the NOAA FPH and the direct tunable diode laser absorption spectrometer (dTDLAS) MC-PicT-1.4 during AquaVIT-2 chamber experiments over 6 days that provides confidence in the accuracy of the FPH measurements. Dual instrument flights with two FPHs or an FPH and a cryogenic frost point hygrometer (CFH) also show good agreement when launched on the same balloon. The results from these comparisons demonstrate the high level of accuracy of the NOAA FPH.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据