Xpert HIV-1 point-of-care test for neonatal diagnosis of HIV in the birth testing programme of a maternity hospital: a field evaluation study

期刊

LANCET HIV
卷 4, 期 10, 页码 E442-E448

出版社

ELSEVIER INC
DOI: 10.1016/S2352-3018(17)30097-8

关键词

-

资金

  1. National Institutes of Health

向作者/读者索取更多资源

Background Point-of-care testing (POCT) among HIV-exposed infants might improve linkage to care relative to laboratory-based testing (LABT). We evaluated HIV-1 POCT at birth in the context of universal LABT in a maternity hospital and describe our implementation experience. Methods We did a field evaluation study between Oct 1, 2014,. and April 30, 2016, at the urban Rahima Moosa Mother and Child Hospital (RMMCH), Johannesburg, South Africa. We aimed to sample consecutive neonates at birth with POCT (Cepheid Xpert HIV-1 Qualitative test) and compared results with those of LABT (Roche COBAS TaqMan HIV-1 Qualitative test) with respect to performance in terms of sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and Cohen's. coefficient, result return, antiretroviral treatment (ART) initiation, and coverage. Findings 18 268 women delivered livebirths at RMMCH and 4267 (23%) were HIV-positive with 4336 HIV-exposed neonates delivered. Mothers of 4141 (96%) HIV-exposed neonates were offered infant birth testing. Mothers of 4112 (99%) neonates consented. In 78 neonates with consent (2%), a test was not done due to early neonatal death (n=13), mother departing before venesection, or staff unavailability. Among 3970 infants who had LABT, 57 (1%) tested positive, 3906 (99%) tested negative, two (<1%) were indeterminate, and five (<1%) had an error result. 2238 (56%) of these infants had concurrent POCT. POCT detected all 30 HIV-infected neonates (sensitivity 100%; 95% CI 88.4-100) with two additional false-positive results (specificity 99.9%; 99.7-100). All positive and 96.2% of negative POCT results were returned compared with 88.9% of positive and 52.8% of negative LABT results. Although every POCT required 90 min of instrument time, 2.6 h (IQR 2.3-3.1) elapsed between phlebotomy and result return. In days, median time of result return for POCT was 1 day, significantly earlier than 10 days for LABT (p<0.0001). ART was initiated in 30 neonates (100%) with positive POCT compared with 24 (88.9%, p=0.10) of 27 infants who had LABT only, with initiation occurring a median of 5 days earlier in the POCT group (p<0.0001). POCT implementation required additional staff and weekend cover. Interpretation Compared with LABT, POCT was associated with good performance, improved rates of result return, and reduced time to ART initiation. Resources needed to integrate POCT into a routine birth testing programme require further evaluation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据